2024, Quality, standardization, metrology, maintenance and exploitation - QSMME, str. 632-635
|
Analysis of results obtained during calibration of force transducers
(naslov ne postoji na srpskom)
Ključne reči: force transducer; static calibration; method
Sažetak
(ne postoji na srpskom)
This paper demonstrates difference between results obtained during calibration of force transducers using standard method, which includes "warm up" of transducer before starting calibration, according to ISO 376 and calibration which does not include "warm up". Transducers were calibrated with dead weight machines Schenck Df-0.2 and DF-1.0. Paper will show is there significant difference between these two methods.
|